Frozen 2026-10-21 04:56 UTC by ESRA (automatic) · published at upload. ← Current dataset page
| calibration range | 0, 5, 10, 25, 30, 50, 75, 100 mM LiCl, each in 4 mL 3 wt% HCl; recalibrated each time the plasma was reignited |
| calibration range source | SI, ICP-AES, p.18 |
| cell id | trial-data |
| digestion method | Not an acid digestion of a solid: liquid wash-solvent sample diluted 1:1 with Milli-Q water, filtered through a 0.22 micron hydrophilic PTFE syringe filter, then 0.4 mL added to 4 mL of 3 wt% HCl |
| digestion method source | SI, ICP-AES, p.18 |
| elements measured | Li, P |
| elements measured source | 240614_150_153.txt column headers (Li610.362nm, Li670.784nm, P213.618nm, P214.914nm, P253.565nm) |
| esra id | ESRA-ICP-0004 |
| experiment type | ICP |
| files public | true |
| institution code | ESRA |
| instrument type | ICP-AES; Agilent 5100 VDV |
| instrument type source | SI, Experimental Methods > ICP-AES, p.18 |
| materials | Celgard 2325 (separator); Cu (working_electrode, current_collector); LP40 (electrolyte); Li (anode, counter_electrode); LiFSI DME (electrolyte); LiFSI DME/TTE LHCE (electrolyte) |
| matrix | 3 wt% HCl |
| matrix source | SI, ICP-AES, p.18 |
| notes | Batch run analyzing many samples (150_01, 150_03-150_13, 153_03-153_06) in one session. None are the same physical cell as this upload's Cycling data (cell 204). Electrolyte is confirmed as LP40 only for 153_05 and 153_06 (cyc_num_metadata.csv); the 150_xx rows have no matching spreadsheet entry, so their electrolyte is not stated anywhere in the attached files. cell_id is left out because this block covers many distinct cells, not one. |
| pi name | Betar M. Gallant |
| project | Energy Storage Research Alliance |
| researcher name | Musanna Galib |
| visibility | public |
| 240614_150_153.txt | 628 B | 2026-10-21 04:56 UTC |
| 240716_154to157.txt | 1.0 kB | 2026-10-21 04:56 UTC |
Size and modification time say whether a file changed since; ESRA does not store checksums.